Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Simpson 270-2,3: (1 year) CAL VER /5500 DATE: 09-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 23 NUMBER OF LINES: 148 CONFIGURATION: Fluke 5500A ============================================================================= # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: NONE 1.004 DISP Temperature: 20 - 50degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DC VOLTAGE} 1.007 DISP Connect the 5500A to the UUT as follows: 1.007 DISP [32] 5500A NORMAL HI to UUT POS 1.007 DISP [32] 5500A NORMAL LO to UUT COMMON 1.008 DISP Set the Function Switch to +DC. 1.008 DISP Set the Range Switch to 1000V. 1.009 HEAD DC VOLTAGE: {1000V Range} 1.010 5500 1000 1000V 2/ 2W 2.001 ASK+ R 2.002 HEAD DC VOLTAGE: {250V Range} 2.003 5500 250 250V 2/ 2W 3.001 HEAD DC VOLTAGE: {50V Range} 3.002 5500 50 50.0V 2/ 2W 4.001 HEAD DC VOLTAGE: {10V Range} 4.002 5500 10 10.0V 2/ 2W 5.001 DISP Set the Function Switch to -DC. 5.002 5500 10 -10.0V 2/ 2W 6.001 DISP Return the Function Switch to +DC. 6.002 HEAD DC VOLTAGE: {2.5V Range} 6.003 5500 2.5 2.50V 2/ 2W 7.001 HEAD {AC VOLTAGE} 7.002 DISP Set the Function Switch to AC. 7.003 HEAD AC VOLTAGE: {2.5V Range} 7.004 5500 2.5 2.50V 3/ 60H SI 2W 8.001 HEAD AC VOLTAGE: {10V Range} 8.002 5500 10 10.0V 3/ 20kH SI 2W 9.001 HEAD AC VOLTAGE: {50V Range} 9.002 5500 50 50.0V 3/ 10kH SI 2W 10.001 HEAD AC VOLTAGE: {250V Range} 10.002 5500 250 250V 3/ 400H SI 2W 11.001 HEAD AC VOLTAGE: {1000V Range} 11.002 5500 1000 1000V 3/ 60H SI 2W 12.001 ASK- R 12.002 HEAD {RESISTANCE} 12.003 DISP Connect the 5500 to the UUT as follows: 12.003 DISP [32] 5500A AUX HI to UUT V/OHM 12.003 DISP [32] 5500A AUX LO to UUT COM 12.003 DISP [32] 5500A NORMAL HI to UUT V/OHM 12.003 DISP [32] 5500A NORMAL LO to UUT COM 12.003 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 12.003 DISP the NORMAL leads to the SENSE leads AT the UUT. 12.004 HEAD RESISTANCE: {Rx1 Range} 12.005 DISP Set the Function Switch to +DC. 12.005 DISP Set the Range Switch to Rx1. 12.006 5500 0.0Z S CW 12.007 DISP Adjust the front panel ZERO OHMS control for a meter 12.007 DISP indication of 0 ohms. 12.007 DISP 12.007 DISP If unit will not zero, replace the D-sized battery. 12.008 5500 10 10.0Z 2% CW 13.001 HEAD RESISTANCE: {Rx100 Range} 13.002 DISP Set the Range Switch to Rx100. 13.003 5500 0.0Z S CW 13.004 DISP Adjust the front panel ZERO OHMS control for a meter 13.004 DISP indication of 0 ohms. 13.004 DISP 13.004 DISP If unit will not zero, replace the D-sized battery. 13.005 5500 1000 1000Z 2% CW 14.001 HEAD RESISTANCE: {Rx10,000 Range} 14.002 DISP Set the Range Switch to Rx10,000. 14.003 5500 0.0Z S CW 14.004 DISP Adjust the front panel ZERO OHMS control for a meter 14.004 DISP indication of 0 ohms. 14.004 DISP 14.004 DISP If unit will not zero, replace the D-sized battery. 14.005 5500 100 100kZ 2% CW 15.001 MESS 15.002 ASK+ R 15.003 HEAD {CURRENT} 15.004 DISP Connect the 5500A to the UUT as follows: 15.004 DISP [32] 5500A AUX HI to UUT POS 15.004 DISP [32] 5500A AUX LO to UUT COMMON 15.005 HEAD CURRENT: {500mA Range} 15.006 5500 500 500mA 2/ 2W 16.001 HEAD CURRENT: {100mA Range} 16.002 5500 100 100mA 2/ 2W 17.001 HEAD CURRENT: {10mA Range} 17.002 5500 10 10.0mA 2/ 2W 18.001 HEAD CURRENT: {1mA Range} 18.002 5500 1 1.00mA 2/ 2W 19.001 HEAD CURRENT: {50uA Range} 19.002 DISP Connect the 5500A to the UUT as follows: 19.002 DISP [32] 5500A AUX HI to UUT 50uA 19.002 DISP [32] 5500A AUX LO to UUT COMMON 19.003 5500 50 50.0uA 2/ 2W 20.001 ASK- R 20.002 HEAD {10 AMP} 20.003 DISP Connect the 5500A to the UUT as follows: 20.003 DISP [32] 5500A AUX HI to UUT +10 AMPS 20.003 DISP [32] 5500A AUX LO to UUT -10 AMPS 20.004 DISP Set the Range Switch to 10mA (AMPS) 20.005 HEAD CURRENT: {10A Range} 20.006 5500 10 10.0A 2/ 2W 21.001 HEAD {HIGH VOLTAGE} 21.002 ASK+ R 21.003 RNG 1000V 21.004 DISP Connect the 5500A to the UUT as follow: 21.004 DISP [32] 5500A NORMAL HI to UUT DC 5000V 21.004 DISP [32] 5500A NORMAL LO to UUT COMMON 21.005 ASK- R 21.006 HEAD HIGH VOLTAGE: {5000V Range} 21.007 5500 5000 1000V 2/ 2W 22.001 DISP Set the Function Switch to AC. 22.002 DISP Connect the 5500A to the UUT as follow: 22.002 DISP [32] 5500A NORMAL HI to UUT AC 5000V 22.002 DISP [32] 5500A NORMAL LO to UUT COMMON 22.003 5500 5000 1000V 2/ 100H SI 2W 23.001 END